Line 168: Line 168:
 ** **
  
-{{:internal:proddocs:3271-12.png|}} {{:internal:proddocs:3271d.jpg|}}+{{:internal:proddocs:3271-12.png|}} 
  
  
Line 176: Line 176:
 ** **
  
-{{:internal:proddocs:3271-13.png|}} {{:internal:proddocs:3271e.jpg|}}+{{:internal:proddocs:3271-13.png|}} {{:internal:proddocs:3271d.jpg?400|}}
  
  
 **To calibrate sender to full, turn rotary switch to the ‘F’ position as indicated in the diagram. **To calibrate sender to full, turn rotary switch to the ‘F’ position as indicated in the diagram.
  
-The sender must be dipped into the tube of fuel to where, the fuel must reach the bottom of the threaded part of the sender body – ensure that the tube has sufficient fuel to reach this level.+The sender must be dipped into the tube of fuel to where, the fuel must reach the bottom of the threaded part of the sender body – ensure that the tube has sufficient fuel to reach this level. Keep clicking the Get Cal button until the Current measured value stops changing. This is a few seconds on a short sender, but could be up to a minute on a 2metre sender. When the value has stopped changing, then calibrate full at the required level. Check that the reading on the big red table goes down very slightly as the sender is raised. DO NOT RESET THE EMPTY, EVEN IF IT READS 2%
  
 Now as for the empty calibration. With a magnet locate the re-calibration place on the top of the unit labelled MAGNET and you will see the blue led remain lit for 2.5 seconds (you can repeat this as many times as you like until you are happy that you have set the FULL position). Now as for the empty calibration. With a magnet locate the re-calibration place on the top of the unit labelled MAGNET and you will see the blue led remain lit for 2.5 seconds (you can repeat this as many times as you like until you are happy that you have set the FULL position).
Line 187: Line 187:
 This is checked with the GET CAL box which should display the FULL values.** This is checked with the GET CAL box which should display the FULL values.**
  
-{{:internal:proddocs:3271-14.png|}}+{{:internal:proddocs:3271-14.png|}} {{:internal:proddocs:3271e.jpg?400|}}
  
  