4291/4925 Build Instructions.

The 4291 PCB assembly contains components that are sensitive to electrostatic discharge (ESD) – please make sure you are wearing your wrist strap when handling.

Programming.

Connect unit to programming device.

  • Select STM32CubeProgrammer
  • Connect to Unit

  • Select Device programming

  • Browse and select the S:/Code/Production/4291/4291bootloader

  * Click Start Programming Making sure all the checked boxes are correct, you should see the following

You can now disconnect the unit, if programming unsucessful, erase the device memory and try again.

  • Open Monex and click 'Get Serial' and 'GetProdInfo', you should see the following

  • Go to the Programming tab and match the device instance to the Target Select

  • Click on 4291 and the Reprogram on the pop-up
  • Select S:/Code/Production/4291/4291

  • The Device should start programming and should 'PASS', and Reboot to Application Code
  • Go Back to Device Info Tab and check that Model ID now shows as 4-20mA Tank Level Adapter

Test Procedure

• There should be no devices that are not mentioned in this document on the CAN bus.

• All tests should be carried out at room temperature unless specified otherwise.

• All tests should be performed fully and in specified order.

• At the start of testing DUT has to be in default configuration state.

===== Calibrating Sender without Hydrostatic Sensor =====

TEST EQUIPMENT USED

  • 4291 Sender
  • Flat Head Screwdriver
  • 4-20mA Test Unit
  • NMEA Network
  • Magnet

Step 1

  • Power up the unit and check connection on Monex
  • Connect the Test Box as shown in the picture below:
  • Begin by calibrating the Sender Empty, do this by setting the Test box to 0 and pressing a magnet to the unit in the shown location.

Calibrating Sender to Sensor

TEST EQUIPMENT USED

  • 4291 Sender
  • Flat Head Screwdriver
  • Hydrostatic Sensor (if calibrating for sensor, otherwise use 4-20mA Test Box)
  • NMEA Network
  • Calibrating Pipe on staircase if calibrating to Hydrostatic Sensor (production side)
  • CTS file if provided by customer

Step 1

Power on the NMEA power supply and set to 12V before turning on.

You will need an NMEA connector in the production staircase, and connect the 4291, the blue LED should be double blinking every few seconds.

Open Monex on your PC, click 'Get Serial' and 'Get ProdInfo', you shoud see the 4291 show up.

For the 4925 tank sender follow Instructions below.

4 Configuration

The device is configured using the left hand rotary switch before use the unit must be calibrated to the sensor.

4.1 Zero calibration

  1. Ensure that the trim sensor is in the zero (full down) position
  2. Turn the 'config' rotary switch to position 'E'
  3. Hold a magnet over the label marking 'Magnet'
  4. The blue LED will light for 2 seconds. The zero position is now calibrated
  5. Return the config switch to the 0 position

4.2 100% Calibration

  1. Ensure that the trim sensor is in the 100% (full up) position
  2. Turn the 'config' rotary switch to position 'F'
  3. Hold a magnet over the label marking 'magnet'
  4. The blue LED will light for 2 seconds. The 100% position is now calibrated
  5. Return the config switch to the 0 position

Step 2

Connect the Hydrostatic sensor to the 4291 as shown:

Step 3

Make sure the water Pipe is empty before connecting the hydrostatic sensor.

Open the Left valve on the staircase pipe and screw in the hydrostatic sensor, make sure it is level to the ground when taking readings and calibrating.

Use the screwdriver and turn the left Hex switch to 'E', place magnet next to 'magnet' label (under both switches), The blue LED will start to blink rapidly, hold the magnet in place until the LED turns solid blue.

Check Monex to see that the sender is transmitting 0%

Step 4

Now proceed to fill the pipe on the staircase to the specified level on the CTS file “full cal”. If no CTS provided, calibrate to the specified maximum height on the sensor (make sure its as precise as possible).

If 4671 then full calibration is at 4 metres.

If 4670, 4672 or 4673 then full calibration is a 2 metres.

When the water level is at the full calibration level make sure the hydrostatic sensor is as horizontal as possible.

Turn the left Hex switch to 'F', place magnet next to 'magnet' label (under both switches), The blue LED will start to blink rapidly, hold the magnet in place until the LED turns solid blue.

Check Monex to see that the sender is transmitting 100%.

Step 5

On monex go into the Volumetric mode tab, click 'Read table from sender'.

Now Drain the water slightly, and each time compare the water level to the monex output, in unsure then ask for help.

If the readings are all within 1% of the real value then calibration is finished.

TEST EQUIPMENT USED

  • • NMEA2000 Network with PCAN-View NMEA2000® USB Adaptor
  • • Test unit
  • • Calibrated Multimeter
  • • MONEX installed on a PC

Fit PCB to box – this will go either way round

Secure with 4 x spacers

Take care when fitting the spacer by the NMEA connector, as it may foul when fitting

Fit top plate and screw down

Ensure the LED’s don’t protrude above the height of the plate – if they are like this they need to be removed and fitted lower down

The screws must also be flush after fitting.

Fit top label

The Adaptor is now ready for Labels, Connect to monex for serial numbers

Manual Printing Instructions

  1. Go to (S) Drive.
  2. Parts Documents.
  3. Select 4291
  4. Select PDF's.
  5. Select 4291 user manual and print.

Label Printing Instructions

  1. Choose labels required
  2. Select (LabelDirect)
  3. Close (Pop Up)
  4. Go to file-open
  5. Select - Parts documents
  6. Select - 4291
  7. Select - Labels
  8. Select - UL_4291_A-2.0.4
  9. Highlight Serial number and change in Text box.
  10. Go to Incrementing number Properties and change serial number.
  11. Go to file -Print Setup
  12. Select bradyBBP11-34L(copy 1) and select Calibrate.
  13. label will calibrate.

Packaging

  • Use RPT103 box
  • 2 parts insert 203613
  • Outer sleeve
  • Contents 1 x 4291 Tank Level Adaptor
  • 1 x User Manual

Manual Printing Instructions

  1. Go to (S) Drive.
  2. Parts Documents.
  3. Select 4291
  4. Select PDF's.
  5. Select 4291 user manual and print.

Label Printing Instructions

  1. Choose labels required
  2. Select (LabelDirect)
  3. Close (Pop Up)
  4. Go to file-open
  5. Select - Parts documents
  6. Select - 4291
  7. Select - Labels
  8. Select - UL_4291_A-2.0.4
  9. Highlight Serial number and change in Text box.
  10. Go to Incrementing number Properties and change serial number.
  11. Go to file -Print Setup
  12. Select bradyBBP11-34L(copy 1) and select Calibrate.
  13. label will calibrate.

For the 4925 the above procedures are the same with different labels and manuals.